Olympus MX61A Manual - Halaman 25

Jelajahi secara online atau unduh pdf Manual untuk Peralatan Laboratorium Olympus MX61A. Olympus MX61A 48 halaman. Motorized 300 mm compatible semiconductor/fpd inspection microscope

13
Setting the DIP Switches
@
Fig. 24
Switch bits (ON: 1, OFF: 0)
DIP switches
1
DIP SW1
0
(Upper switch)
1
SW2
0
(Lower switch)
}The masked fields indicate the factory setups (all OFF).
}The following table shows the details of DIP switch settings.
# Set the main switch @ of the microscope to "
the DIP switch.
²
This is because the DIP switch settings are read and entered only at
the moment the microscope is turned ON.
³
# Before setting the DIP switches, discharge the static electricity from
your body by lightly touching any metallic object.
1. When using the U-D6REMC or U-D6BDREMC revolving nosepiece, be
sure to set DIP SW1 bit 3 ² to "ON".
When using the U-D5BDREMC or U-P5BDREMC revolving nosepiece,
leave the DIP SW1 bit 3 in the default setting of "OFF".
2. To inhibit initialization at the moment the microscope is turned ON, be
sure to set DIP SW1 bit 7 ³ to "ON".
2
3
4
5
6
7
8
0
0
0
1
1
0
1
1
0
0
0
1
0
1
0
0
0
0
0
0
0
Function
Buzzer inhibition
Revolving nosepiece type
Reserved for maker
Reserved for maker
Freedom
(number of positions) search
Power-ON initialization inhibition Initialize
Reserved for maker
(Fig. 24)
" (OFF) before setting
Setting Result
Generate buzzer
Do not generate
U-D5BDREMC/ U-P5BDREMC
U-D6REMC/ U-D6BDREMC
Not used
Not used
Fixed at OFF
Fixed at OFF
Search during power ON initialization
Do not search
Do not initialize
All fixed at OFF
MX61A
22