Olympus MX61A Manual - Halaman 37

Jelajahi secara online atau unduh pdf Manual untuk Peralatan Laboratorium Olympus MX61A. Olympus MX61A 48 halaman. Motorized 300 mm compatible semiconductor/fpd inspection microscope

@
|
CONFIG 4/6 screen
²
PFCL ON/OFF selection screen
6

CONFIG 4/6 Screen Setup

}The CONFIG 4/6 screen is used to set the parfocal correction values of
the objectives.
The default values are "0" for all objectives.
1. Touch the objective No. switch @ for the objective with the lowest magni-
³
fication to engage it in the light path.
2. Place a specimen on the stage and adjust the focus.
}The operations in steps 1 and 2 above are intended to facilitate focus
adjustment using the objective with the highest magnification.
3. Touch the objective No. switch @ for the objective with the highest mag-
nification, and bring the specimen into focus with the new objective.
4. Touch the PREFOCAL SET-UP switch ².
5. Touch the objective No. switch @ for the objective with the second high-
est magnification.
The current Z-stage position is displayed in the "OFFSET" field for the
objective with the highest magnification.
6. Bring the specimen into focus.
7. Follow steps 5 and 6 to set up the objectives with the third and fourth
highest magnifications.
8. Touch the SET switch ³. The OFFSET value with reference to the objec-
tive with the highest magnification (assumed to be "0") is displayed.
Other Switches
|PFCL: Touch the
which you can select whether the parfocal correction is applied
(ON) or not (OFF).
ƒ
After touching ON or OFF, touch the OK switch ƒ to enter the
selection.
...
...BACK: Touch this switch to return to the CONFIG 3/6 screen without
applying the parfocal correction value setup.
†NEXT: Touch this switch to advance to the CONFIG 5/6 screen.
The parfocal correction value set in the CONFIG 4/6 screen
starts to be applied from the CONFIG 5/6 screen.
switch to open the PFCL selection screen in
MX61A
34