Olympus U-RFLT50 Petunjuk Manual - Halaman 12

Jelajahi secara online atau unduh pdf Petunjuk Manual untuk Peralatan Laboratorium Olympus U-RFLT50. Olympus U-RFLT50 16 halaman. Power supply unit
Juga untuk Olympus U-RFLT50: Petunjuk Manual (36 halaman)

Olympus U-RFLT50 Petunjuk Manual
5
5

TROUBLESHOOTING GUIDE

Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If problems occur,
please review the following list and take remedial action as appropriate. If you cannot solve the problem after checking the
entire list, please contact your local Olympus representative for assistance.
Problem
1. The main switch indicator does not
light up.
2. Main switch indicator lights, but the
mercury burner does not ignite.
3. The burner flickers or is dark.
4. The burner life is exceptionally short. The input voltage selector switch is set
7
Cause
Improper connection of the power cord.
Fuses are blown.
Improperly connected connectors.
The burner is not installed.
The lamp housing interlock mechanism
is activated.
Auto ignition is malfunctioning.
The auto ignition interlock mechanism is
activated.
Insufficient time has elapsed since the
burner was turned on.
The voltage selector switch and/or frequen-
cy selector switch is set incorrectly.
The burner life has expired.
The input voltage selector switch is set
incorrectly.
incorrectly.
Remedy
Connect the power cord correctly.
Replace fuses.
Connect correctly.
Install the burner.
Tighten the burner socket clamping screw.
Turn OFF the main switch on the power
supply unit. Turn ON again. (Repeat as
necessary.)
Press the switch inside the starter reset
hole on the right side of the power supply
unit, and then press the main switch to " I "
(ON) again.
Wait for at least 10 minutes after igniting
the burner.
Set switches to conform with the local AC
mains.
Replace the mercury burner when the
burner life time hour counter reading
exceeds 100 hours. (Burner service life is
100 hours.)
Set to conform with the local AC mains.
Set to conform with the local AC mains.