c. The inlet assembly can now be disassembled by using a small flathead
screwdriver to unscrew the inlet valve (VRA-112) from the vent plug (VRA-111).
Be careful when disassembling as the inlet assembly is under spring tension and
small parts could be lost.
d. Once the inlet valve, vent plug, spring, spring holder and spring retainer are
removed, the seal plug (VRA-182) and inlet capsule (VRA-141) can be
unscrewed.
e. The inlet valve seat (VRA-110) can be removed by pressing it up and through
the inlet capsule. The inlet valve seat is a maintenance part and once removed
should never be reused.
f. Carefully clean the inlet capsule, seal plug, inlet valve stem, vent plug, spring
holder, spring retainer and spring. NOTE: The VRA-112 inlet valve stem must be
cleaned using a non-abrasive cloth (or Scotch-Brite pad).
g. Using a new inlet valve seat and new o-rings (with a thin film of the grease
provided on each o-ring), reassemble the inlet assembly in reverse order.
-Should you have any questions during maintenance of your VR3 vacuum regulator, please
contact your local service provider or Archer Instruments for support.