Accuris W3100-210 Panduan Pengoperasian - Halaman 8
Jelajahi secara online atau unduh pdf Panduan Pengoperasian untuk Timbangan Accuris W3100-210. Accuris W3100-210 16 halaman. Analytical
Juga untuk Accuris W3100-210: Panduan Pengoperasian (16 halaman)
Control Panel and Display:
On/Off
Print
Tare
4. CALIBRATION
Background
The Accuris analytical balance uses the equilibrium principle of
electronic magnetic force. Among the various factors that may affect the
accuracy of the balance, gravitational force is the most significant. In
different geographical areas on the earth, gravitational force varies, and a
balance should be calibrated at its specific installation location.
The Accuris Analytical Balance can be calibrated using the internal
"Quick-Cal" feature (only available on models W3100A-120 and
W3100A-210) or by calibrating to a known reference weight.
TM
4-1 INTERNAL "QUICK-CAL
" (Models W3100A-120 and
W3100A-210)
In this mode, FUNCTION C1 should be set to 0 (C1- - 0). See section 5,
Function Setting.
8