1. When the balance is in Standby Mode (plugged in to power,
On/Off set to Off) press the CAL button.
2. The display will show 1. Press TARE then CAL, TARE, CAL
alternatively and the number on the display will increase: 1, 2, 3, 4,
etc.
3. When the display shows 8, press the On/Off button. The display
will show 8888888
4. Press CAL to enter linear calibration mode, and wait for the
balance to carry out the zero point calibration and Lnr 50 will be
displayed
5. When Lnr 50 is displayed, place a 50g calibration weight on the
pan and press TARE. Lnr will be displayed and then Lnr 100.
6. When Lnr 100 is displayed, place a 100g calibration weight on the
pan and press TARE. Lnr will be displayed and then Lnr 150.
7. When Lnr 150 is displayed, place a 150g calibration weight on the
pan and press TARE. When linear calibration is complete the
display will show 0.0000g.
8. Remove the calibration weight and press TARE. Linear calibration
is complete.

4-5 200g CALIBRATION

The W3100A-210 balances may require an additional calibration near the
maximum capacity of the balance (200g Calibration). If after calibration
and linearity adjustment, the values are off in the range of 180g to 200g,
a 200g calibration is required.
Follow these steps:
1. Switch on the balance, so "0.0000g" is displayed
2. Place a suitable 200g calibration weight in the center of the weighing
pan and close the draft shield drawer.
3. Press SET and " ---------------" is displayed.
4. Press MODE key, and "CAL 200" is displayed.
11