SCANREADER OPTION INTERFACE SETTINGS

The ScanReader option interface is factory set and
normally requires no changes. There is no user
interface menu, but changes can be made using
the method described in Changing Selections on
page 4. The notes below define the permitted data
entry values for the various parameters. The
ScanReader options are at security access level 4.
325 Serial Number 4
10 character string. Allowed characters are the
digits 0 to 9 and alphabetics a to z.
Upper case will be transmitted as lower case.
Default setting is 0000000000
This is the serial number that is reported via the
ScanReader interface.
If this is set to 0000000000 the rightmost 9 digits
of the sensor ID number prefixed with a 7 will be
used as the ScanReader serial number.
Invalid characters will be ignored.
326 Adapter Code 4
0 = ScanCoder
1 = AquaMaster
Default setting is 0
Selects the error code capability of the interface.
'0' will provide no error code information.
'1' will provide extended error code information to
enabled Readers.
8
327 Totaliser Source and Range 4
0 = Fwd
x1
1 = Rev
x1
2 = Net
x1
3 = Fwd
x10
4 = Rev
x10
5 = Net
x10
6 = Fwd
x100
7 = Rev
x100
8 = Net
x100
Default setting is 0
Selects the Totaliser source and resolution range
for the Reader value field. Negative totals will be
reported as 0.
ScanReader interface only has 6 digits defined for
Totaliser value. Range setting defines how the 6
digits are used. For example, for range x100, a
total of 12345678 would be conveyed to the
Reader as 123456.