Accusom Home Sleep Test Panduan Pengguna - Halaman 7
Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Produk Perawatan Pribadi Accusom Home Sleep Test. Accusom Home Sleep Test 8 halaman. Test
Juga untuk Accusom Home Sleep Test: Panduan Pengguna (9 halaman)
Frequently Asked Questions
What if I need to get up in the middle of the night?
There is nothing special to do if you need to get up at night. Leave
the sensors in place and connected to the AccuSom on your arm.
If any of the sensor loses connectivity with the AccuSom device, a
voice prompt will indicate which sensor needs to be repositioned or
plugged back into the device.
What if the voice prompt says "Check Chest Sensor"?
If the voice prompt says "Check Chest Sensor": Reposition and
tighten the Chest Sensor. The sensor should be placed just below
the breasts. Once the Chest Sensor has been correctly repositioned
and tightened, and the connection to the AccuSom device has been
verified, the voice prompt will say "Chest Sensor OK" and Chest Sen-
sor light will turn green. This may take about 30 seconds. If the Chest
Sensor LED light is still not green, squeeze the Chest Sensor Belt
tubing between two fingers. If the LED light still does not turn green,
call NovaSom, Inc. at 1-877-753-3776.
What if the voice prompt says "Check Breath Sensor"?
If the voice prompt says "Check Breath Sensor", please reposition the
Breath Sensor so that it is resting just above your upper lip and is
centered below your nose. Make sure the side with the five small holes
is facing your mouth. Once the sensor has been correctly positioned,
and the connection from the Breath Sensor to the AccuSom has been
verified, the voice prompt will say "Breath Sensor OK" and the Breath
Sensor light will turn green. This may take about 30 seconds. If the
LED light still does not turn green, call NovaSom, Inc. at
1-877-753-3776. If you are a very active sleeper, this may happen
several times during the night.
What if the voice prompt says "Check Finger Sensor"?
If the voice prompt says "Check Finger Sensor", check to make sure
the Finger Sensor is correctly positioned on the fingertip and that
the cable is connected to the device and that you have completely
removed any dark nail polish. Once the Finger Sensor has been
correctly repositioned, the Finger Sensor light will turn green, and the
voice prompt will say "Finger Sensor OK". This may take about 30
seconds. If the LED light still does not turn green, call NovaSom, Inc.
at 1-877-753-3776.
(Frequently asked questions are continued on back cover.)