Section 5: Modbus Registers
Address Variable
16 Bit Int registers
40000 Airflow Angle
40001 Direct airflow velocity x100
40002 Linear airflow velocity x100
40003 Linear airflow velocity x100
40004 Linear airflow velocity x1
40005 Linear airflow volume x100
40006 Linear airflow volume x1
40007 Linear airflow volume x100
40008 Heartbeat register
32 Bit float registers
40010 Airflow Angle
40012 Direct airflow velocity raw
40014 Linear airflow velocity
40016 Linear airflow velocity
40018 Linear airflow velocity
40020 Linear airflow volume
40022 Linear airflow volume
40024 Linear airflow volume
User Settings - holding registers (16bit ints)
40100 K factor (x100)
40101 area in meters (x100)
40102 Modbus Address
40103 Baud Rate
40104 Parity
40105 stop bits
Units
Description
°
actual value from single point
m/s
actual value from single point
m/s
calculated from angle: airflow*cos(angle)
ft/s
calculated from angle
ft/min
calculated from angle
calculated from angle:
m3/s
(airflow*cos(angle))*area
CFM
calculated from angle
KCFM
calculated from angle
Counter increases by 1 approx. every second
°
actual value from single point
m/s
actual value from single point
m/s
calculated
ft/s
calculated
ft/min
calculated
m3/s
calculated
CFM
calculated
KCFM
calculated
1 to 255
default 192
0 = 4800
default 9600
1 = 9600
2 = 19200
3 = 115200
0 = none
default none
1 = odd
2 = even
0 = 1 stop bit
default one
1 = 2 stop bit
9