I. MAIN MANIFOLD
1. Take main manifold from the shipping carton and separate manifold cover from the main mani-
fold block.
2. Remove the brass caps from both the nitrous oxide gas line (left) and the oxygen gas line (right)
that are located in the tank room.
3. Remove the red caps from the outlet fittings at the top of the main manifold and attach the main
manifold to the gas lines using the DISS fittings.
Note: Use 2 wrenches when tightening the fittings.
4. Using the three 1/2-inch screws, secure the main manifold to the wall through the mounting holes
in the main manifold block. The manifold must be installed a minimum of five feet from the floor.
Note: When studs are not available to secure the manifold block directly to the wall, a
mounting board 18 inches wide, 4 inches tall and 1 inch thick (secured to the wall studs)
is recommended.
5. Remove the red cap from the light gray cable and plug the cable into the connector (cable port) at
the top of the main manifold circuit board. Push the cable connector into the port until there is a
click and the cable connector cannot be pulled out of the port.
PRESSURE
N
O
2
Line Pressure
INSTALLATION
Guardian Monitor™
M A N I F O L D
PRESSURE
O
2
Line Pressure
Use No Oil
P.S.I.
PRESSURE
4
P.S.I.
PRESSURE