Replacing the horizontal lift brake
Preliminary steps
CAUTION
2
1
Fig. 1
•
Connect the SIREMOBIL Iso-C to power .
•
Position the wooden block (1/Fig. 1) underneath.
•
Move the C-arm into the horizontal position (Fig. 1).
•
Position the C-arm over the work table.
•
Slowly move the lifting column down in increments until the C-arm is positioned just flat
on the table. When doing this, make sure that the work table is centered around the C-
arm and that as much of the railing as possible is lying on the table.
NOTICE
Disassembly
•
Switch OFF the SIREMOBIL Iso-C.
•
Remove the docking plate for the MODULARIS, if it is connected.
•
Remove the cover with the brake switches (2/Fig.1) (screws are on the underside).
When doing this, remove the ground cable and disconnect the connection to the switch
board.
Siemens AG
Medical Engineering
Push the wooden block under the SIREMOBIL frame prior to removing
the C-arm for service. Otherwise, the frame may tip over when the C-arm
is removed from the lifting column.
When placing the C-arm on the table, lower the lifting column only far
enough until the C-arm is positioned flat on the table with the entire rail-
ing.
Fig. 1 shows one possible arrangement using 2 smaller work tables.