WARNING
CAUTION
SIREMOBIL Iso-C
Risk of injury from mechanical parts! If not observed, minor to
more severe injury, especially to the hands, can occur.
The C-arm of the system is manufactured from carbon fiber com-
posite material (CFK). In the case of damage to the carbon fiber
structure, for example caused by collision or by cutting or scrap-
ing tools, individual carbon fibers or carbon fiber bundles can
stand out and on contact cause injuries to the skin or penetrate
into the tissue of the hand.
Before starting work on the C-arm, make sure that there
is no damage to the carbon fiber structures (visual
inspection).
If smaller superficial damage is detectable in the painted
area or in the area of the running surface edges of the
C-arm, wear working gloves and remove any projecting
carbon fibers. Then carefully rub the damaged places flat
with a fine emery cloth and close the damaged surface
with paint. A spray paint in the color of the C-arm can be
used for this purpose (see the Service Tool Catalog).
Remove any carbon fibers lying in the C-arm profile with
a damp sponge or cloth.
If greater damage to the carbon fiber structure indicating
general or large-scale destruction of the carbon fiber
structure (e.g., crack formation or flaked-off places on
the running surfaces with torn carbon fibers) can be
detected, replace the C-arm.
Risk of injury from mechanical parts! If not observed, minor to
more severe injury, especially to the hands, can occur.
Parts such as flat plugs, threaded bolts, cut-off cable ties and com-
ponent edges that, if care is not taken, can cause crushing, abra-
sion and cuts to the skin, particularly to the hands, can be touched
after the covers are opened.
Perform the required work with special care and atten-
tion to detail.
If needed, wear working gloves.