Inspection and maintenance
QSQ
Central beam
Check the coincidence of the radiation field and film center and, if necessary, adjust it
according to Adjustment Instructions RLL5-310.071.01..., chapter 10, page 10-2.
2.7.3

Collimator blades

PMA
Belt tension
Measure the tension of the drive belt for the primary collimator blades:
Nominal value = 5 N at 22 mm.
If necessary, correct the tension by adjusting the drive motor.
2.7.4

Cassette drive

PMA
Belt tension
Check the tension of the drive belt for the cassette drive according to Adjustment
Instructions RLL5-310.071.01., chapter 8, page 8-1 and, if necessary, adjust it.
NOTICE
Siemens AG
Medical Engineering
If the tension was adjusted, reference switch S51 for the cassette
height measurement must be recalibrated. Refer to Adjustment
instructions RLL5-310.071.01.., chapter 8, page 8-3.