Siemens SITRANS L Pointek ULS200 Panduan Petunjuk Pengoperasian - Halaman 11

Jelajahi secara online atau unduh pdf Panduan Petunjuk Pengoperasian untuk Beralih Siemens SITRANS L Pointek ULS200. Siemens SITRANS L Pointek ULS200 47 halaman. Ultrasonic non-contacting switch

7. The certification of this equipment relies upon the following materials used in its
construction:
Aluminum alloy T356 T6 (main enclosure) and A356 T6 (lid)
GE Lexan 943A polycarbonate
Two-part epoxy encapsulant
Silicon based coating
Santoprene 111-55 gasket
Master Bond Polysulphide EP21LPT or Dow Corning 3-4207 encapsulant
(transducer)
ETFE (transducer)
Epoxy syntactic foam (transducer)
If the equipment is likely to come in contact with aggressive substances, then it is the
responsibility of the user to take suitable precautions that prevent it from being adversely
affected, thus ensuring that the type of protection is not compromised.
8. Equipment Marking:
The equipment marking contains at least the information on the product nameplate,
shown on the inside front cover of this manual.
9. Special Condition for Safe Use: The apparatus must only be supplied from a circuit
containing a suitable rate fuse having a breaking capacity of at least 4 000 A.
Pointek ULS200
Operating Instructions, 01/2019, A5E32268616-AB
Aggressive substances:
Suitable precautions:
e.g. acidic liquids or gases that may attack
metals, or solvents that may affect polymeric
materials
e.g. regular checks as part of routine inspec-
tions or establishing from the material's data
sheet that it is resistant to specific chemicals.
11