- Before using the NEW ASKIR 118 BASIC, consult the instructions for use: failure to read all the instructions in this
manual can be harmful for the patient.
- The device cannot be used to drain chest fluids;
- The device must not be used for suction of explosive, corrosive or easily flammable liquids.
- NEW ASKIR 118 BASIC is not suitable for MRI. Do not introduce the device in MRI environments.
IMPORTANT INFORMATION FOR CORRECT DISPOSAL OF THE PRODUCT IN ACCORDANCE WITH EC
DIRECTIVE 2012/19/UE-RAEE: The symbol on the device indicates the separated collection of electric and
electronic equipment. At the end of life of the device, don't dispose it as mixed solid municipal waste, but
dispose it referring to a specific collection centre located in your area or returring it to the distributor, when
buying a new device of the sample type to be used with the same functions. This procedure of separated
collection of electric and electronic devices is carried out forecasting a European environmental policy aiming
at safeguarding protecting and improving environment quality, as well as avoiding potential effects on human
health due to the presence of hazardous substances in such equipment or to an improper use of the same or of
parts of the same. Caution: The wrong disposal of electric and electronic equipment may involve sanctions.
Model
Typology (MDD 93/42/EEC)
UNI EN ISO 10079-1 Classification
Power Feeding
Maximum Suction Pressure (adjustable)
Minimum Suction Pressure (adjustable)
Maximum Suction Flow
Weight
Insulation Class (when used with the ASKIR 118 SUPPORT
BRACKET mod. ASU 118)
Insulation Class (when used with an Internal battery)
Size
Battery Holding Time
Battery Time Charge
Noise Level
Working Condition
Conservation condition and Transport
Please note that the maximum suction value can vary according to reduced atmospheric pressure if the device is used at high
altitude. The operator must take use at high altitude into account. In such conditions, the vacuum produced by the internal pump
can considerably drop due to the reduction of atmospheric pressure.
Use a soft dry cloth with not – abrasive and not – solvent detergents. To clean the device external parts always use a cotton cloth
dampened with detergent. Don't use abrasive or solvent detergents.
PARTICULAR CARE SHOULD BE TAKEN TO ENSURE THAT THE INTERNAL PARTS OF THE EQUIPMENT DO
NOT GET IN TOUCH WITH LIQUIDS. NEVER CLEAN THE EQUIPMENT WITH WATER.
During all clearing operations use protection gloves and apron (if need be, also wear a face mask and glasses) to avoid getting in
contact with contaminating substances (after each utilization cycle of the machine).
CONTRAINDICATIONS

TECHNICAL CHARACTERISTICS

12V
Room temperature:
Room humidity percentage:
Atmospheric pressure:
Room temperature (≤ 1 mounth): :
Room Temperature (≤ 3 mounth):
Room Temperature (≤ 1 year):
Room humidity percentage:
Atmospheric pressure:

CLEANING DEVICE

3
NEW ASKIR 118 BASIC (REF RE 410170)
Class IIa Medical device
HIGH VACUUM / HIGH FLOW
5,2 A - 14,8 V
by internal Li-Ion battery
4A by ambulance adapter model ASKIR 118
SUPPORT (REF ASU 118)
-75kPa (- 0.75 bar)
-15kPa (-0.15 bar)
36 l /min
2.50 Kg
Class II
Internally Powered Equipment
350 x 150 x 190 mm
70 minutes
360 minutes
Without Suction
54,7 dB
During Maximum Suction
68,2 dB
5 ÷ 40°C
0 ÷ 85% RH
800 ÷ 1060 hPa
- 20°C ÷ 45°C
- 20°C ÷ 35°C
0°C ÷ 25°C
0 ÷ 85% RH
500 ÷ 1060 hPa