Argon Medical Devices SKATER Petunjuk Penggunaan Manual

Jelajahi secara online atau unduh pdf Petunjuk Penggunaan Manual untuk Peralatan Medis Argon Medical Devices SKATER. Argon Medical Devices SKATER 12 halaman. Introducer
Juga untuk Argon Medical Devices SKATER: Panduan Memulai Cepat (2 halaman), Manual (13 halaman)

Argon Medical Devices SKATER Petunjuk Penggunaan Manual
en
Instruction For Use
bg
Инструкции за употреба
cs
Pokyny pro použití
da
Brugsanvisning
de
Gebrauchsanweisung
Οδηγίες χρήσης
el
es
Instrucciones de uso
et
Kasutusjuhised
fi
Käyttöohjeet
fr
Mode d'emploi
hu
Használati utasítás
it
Istruzioni per l'uso
lt
Naudojimo instrukcijos
lv
Lietošanas instrukcija
nl
Gebruiksaanwijzing
no
Bruksanvisning
pl
Instrukcja obsługi
pt
Instruções de utilização
ro
Instrucţiuni de utilizare
sk
Návod na použitie
sv
Bruksanvisning
tr
Kullanım talimatları
ARGON MEDICAL DEVICES, INC.
1445 Flat Creek Road
Athens, Texas 75751 USA
Tel: 800-927-4669
Tel: +1 903-675-9321
www.argonmedical.com
EMERGO EUROPE
Prinsessegracht 20
2514 AP The Hague
The Netherlands
The symbols glossary is located electronically at
www.argonmedical.com/symbols
0086
IFU7511/0617B
SKATER™ Drainage Catheter Non-Locking Pigtail
en - English
Indication for use
The product is for drainage of cysts, abscesses,
haematomas, pleural exudates, ascites, gall bladders
and nephrostomies.
Duration
Up to 12 weeks
Warnings
This device was designed, tested and manufactured
for single use only. Reuse or reprocessing has not
been evaluated and may lead to its failure and
subsequent patient illness, infection or other
injury. Do not reuse, reprocess or re-sterilize this
device.
Do not use if package appear open or if the expiry
date has been exceeded.
The product must only be used by qualified
personnel who are familiar with the technique.
Preparation
Advance the straightener carefully over the
curvature of the catheter, while straightening the
curvature with your fingers.
Introduce the chosen stiffener (if plastic: activate the
hydrophilic coating with saline) and fasten the
stiffener onto the catheter (Luer Lock).
Remove the straightener
Activate the coating by saline.
Procedure
Advance the catheter over the guidewire that has
been introduced into the patient using standard
technique.
When the catheter tip is inside the cavity, loosen the
stiffener.
Hold the stiffener tight and advance the catheter
over the stiffener until the complete length of the
pigtail is in the cavity.
Remove the stiffener.
Verify correct position of the catheter using
fluoroscopy.
Remove the guidewire
Connect the catheter to a drainage bag using a
suitable connecting tube.
It is recommended affixing the catheter in a straight line
and that any curvature be applied to the connecting tube.
It is recommended that the physician follow the
hospital's standard of care procedures for drainage
catheters.
Removal of catheter
Disconnect the drainage bag connector tube from
catheter.
Pull the catheter out gently. If access is to be
maintained, a straight floppy tip guidewire passed
through the catheter will facilitate removal while
maintaining access.
1 / 12