CURRENT PROBE - MN02 DRAWING
ELECTRICAL SPECIFICATIONS
Nominal Range: 100 A
Measurement Range:
50 mA to 100 A (1 Ω load)
50 mA to 90 A (10 Ω load)
Transformation Ratio: 1000:1
Output Signal: 1 mA/A from (1 to 10) Ω
Accuracy and Phase Shift*:
Accuracy 1 Ω load: ≤ 1 % Reading ± 0.02 A
Accuracy 10 Ω load: ≤1.5 % Reading ± 0.01 A
Phase Shift: < 3 ° with 1 Ω load
< 6 ° with 10 Ω load
*Reference conditions: (18 to 28) °C,
(20 to 75) % RH, external magnetic field
< 40 A/m, (48 to 65) Hz sine wave, distor-
tion factor less than 1 %, no DC compo-
nent, no external current carrying conduc-
tor, test sample centered.
Load impedance 1 Ω or 10 Ω.
Overload: 170 A for 10 min ON, 30 min OFF
Frequency Range: (48 to 10,000) Hz
Open Secondary Voltage: ≤ 30 V
Working Voltage: 600 Vrms
Common Mode Voltage: 600 Vrms