Device Information
Mode
Key
Function
PEAK HOLD
Scroll down the list of information fields, i.e. the
next field in the list.
TEST
Scroll up the list of information fields, i.e. the
previous field in the list
UNITS
Show the label of the current information field.
F+CLR/ZERO
Return to normal operating mode.
Label
Typical Display
SErno
Eltno
SoFt
rAnGE
C-dAt
C-rEF
C-cnt
OLOAd
Addr
8
To acitvate the device information mode from normal operating mode use one
hand to press and hold the F key and use the other hand to press the UNITS
key. Return to normal operations by using one hand to press and hold the F
key and use the other hand to press the CLR/ZERO key.
Upon entering device information mode, the selected unit annunciator will
flash (indicating that the device is out of the normal operating mode) and the
display will show the device serial number. Use the keys described in Table 2
to view the information fields described in Table 3.
Table 2
Key Function in Device Information Mode
Table 3 contains the information fields and typical values you can view in the
device information mode. The fields appear in the sequence shown. Scrolling
past the last item in the list will return the display to the first item in the list.
Description
512056
The device serial number (6 digits).
201360
Load element number.
2.60
The software release number.
20
Lb
Device range in 1,000s of pounds.
1
96.08.02
Year, Month and Day of last calibration.
100123
Reference to calibration document number (6 digits).
3
Total number of calibrations performed.
0
Number of times the device has been overloaded beyond 125% of full
capacity.
100
The communication address assigned to the device in multidrop
applications.
Table 3
Device Information Fields
Display
The "label" for the field is shown briefly
when the key is kept depressed. The
actual information value is then shown.
Same as for PEAK HOLD
Same as for PEAK HOLD