MSI MSI-8000 Panduan Operator - Halaman 29
Jelajahi secara online atau unduh pdf Panduan Operator untuk Monitor MSI MSI-8000. MSI MSI-8000 44 halaman. Rf remote display

View Summed Channels
1) Program a function key to "ttl.rd"
(Total Remote Devices).
example we'll use
2) The current channel is displayed.
Press
F2 ttl.rd
display to the summed total.
3) The display briefly displays
"AddAll". Then summed total
weight is displayed.
4) Subsequent presses of the
F2 ttl.rd
key will show all enabled
sum types.
5.5 Behavior Of Zero And Tare In Multiple Channel Systems
Whatever channel that is displayed is considered the "Focus Channel." Pressing the
or
keys will only affect the displayed channel. When displaying summed channels,
TARE
zero and TARE commands will be sent to all devices that contribute to the displayed
weight. For example, if in the "BOTH" mode, and you are displaying Pair 1 (the sum of
SC0 and SC1), pressing
using the ALL mode, then pressing
Using the TARE function: If one device is tared in the individual display mode, the
summed weight will be the sum of a Net weight and a Gross weight. If the TARE key is
pressed when displaying any of the summed modes, all devices that add to form the
current display are Tared and placed in Net mode.
5.6 Printer Set Up
The RS 232 Comm Port is capable of outputting tension data. All the weight modes the RF
linked weight device can measure are available in user formatted form. The control Mode
programs are what cause the MSI-8000 to print.
The modes are:
1. Assigned F-Key is pressed, one transmission of the selected string type is output.
2. On Load - When the tension on the link is stable, one transmission will output. Then
the tension must return to zero to enable another print output.
3. Continuous - Program the interval in seconds up to 65,535 seconds. Setting the
interval to 0 will set an interval as fast as the system can go. To disable printing set
the control mode to "OFF."
The Comm Port settings are independent of any print settings in connected scales. They
reside only in the MSI-8000.
In this
F2
for "ttl.rd".
to change the
Figure 5-5. View Summed Channels
will zero only SC0 and SC1. If displaying the grand total
ZERO
See Function Key setup
on page 15.
F2
TOTAL REM
DEVICES
MOMENTARY
F2
TOTAL REM
DEVICES
MOMENTARY
F2
TOTAL REM
DEVICES
MOMENTARY
will zero all connected sensors.
ZERO
M
Ton
Ton
M
kg
kN
lb
kg
kN
lb
M
Ton
kg
kN
lb
M
Ton
kg
kN
lb
ZERO
COMM Set Up
25