6. How to adjust "flow detecting setting"
a) Press the start button and the unit will display the total number of tests
performed for a second and will begin a countdown.
b) During the countdown, press and hold the mode button for more than 5
seconds until the unit displays
Press mode button shortly until the unit displays
c) The flow detecting setting can be adjusted by pressing the start button.
d) Three kinds of settings are available - Low, High, User.
Flow Detection Settings
Use this option to adjust how hard a user is required to blow.
Press Start to change the settings.
Low level flow detection
This is the default setting for the unit.
High level flow detection
Press start to change to the high level flow setting.
This level requires a stronger breath by the user and is
recommended if previous test attempts have been unsuccessful.
User activated test mode
Press Start again to change to the USER
activated test mode. When testing in this
mode, the operator must press Start to take a breath
sample during blowing and to perform a test.