CALIBRATION
Calibration of the AlcoHAWK
PT400 occurs during manufacturing using
®
advanced alcohol simulation equipment. The manufacturer passes
known alcohol concentrations through the sensor to set baseline values
for testing. During testing, the unit compares users' breath samples to
these baseline values. If the verification determines that the BAC is not
consistent with actual BAC levels, a technician calibrates the unit. The
accuracy of breath alcohol testers can fluctuate over time, depending on
frequency of use and operating conditions.
If the unit is providing inconsistent results or unusually high or low
readings, the unit may need calibration. A general rule of thumb is to
have the unit calibrated every 200 tests (see Test Counter, pg. 8).
However, at a minimum, a technician should recalibrate this unit
at least every six to twelve months. Monthly calibrations may be
required if the unit is used daily.
Please visit www.q3i.com or call 888-399-1687 for calibration or support.
ONE-YEAR LIMITED WARRANTY
Q3 Innovations (Q3I) warrants this unit to be free from defects in
workmanship or material (excluding calibration) under normal use for
ONE (1) YEAR from the date of retail purchase by the original buyer
("Warranty Period").
The warranty covers parts and labor on covered repairs and
ground freight back to the customer (USA only). Except as pro-
vided above, Q3I provides this product "as is."
If a defect arises and Q3I receives a valid claim within the War-
ranty Period, at its option, Q3I will either (1) repair the product,
(2) exchange the product with a product that is new or which has
been built from new or serviceable used parts and functionally
equivalent to the original product, or (3) refund the purchase price
of the product. If a refund is given, the returned product becomes
10