ALCOSIM SIMULATOR SETUP
CAUTION! Ensure that all equipment is unplugged before assembling or
disassembling. Failure to do so may result in damage to the heater element and
risk of electric shock.
Visually inspect the ALCOSIM simulator top housing and container to verify that they
are clean and free of any cracks.
1.
Unscrew the top housing and remove by lifting it up.
2.
Set the top housing aside and fill the solution container with a
new bottle of 20 mg% alcohol reference solution. Do not fill
above or below the 500 mL fill line.
3.
Place the top housing back onto the solution container and
tighten securely. Do not over tighten as this may crack the
container.
4.
Connect the vinyl tubing to the AIR INLET TUBE on the simulator. Connect a
round mouthpiece to the tubing from the AIR INLET TUBE.
5.
Check for leaks in the seal by blowing into the round mouthpiece and covering the
VAPOUR OUTLET TUBE. Air should not bubble through the solution. In case of
bubbling, refer to the "Troubleshooting" section of this manual.
6.
Connect a diffuser to the VAPOUR OUTLET TUBE.
7.
Plug in the simulator power cable and switch the simulator on. The mixing
propeller starts to rotate.
8.
Set the ALCOSIM simulator aside to warm up.
NOTE:
•
It takes approximately 15 minutes for the solution to reach 34 °C. The current
temperature is displayed on the LCD display
•
Replace the alcohol reference solution every 20 tests / 5 days to maintain
proper alcohol concentration and accuracy