Alcolock V3 Panduan Kalibrasi - Halaman 3

Jelajahi secara online atau unduh pdf Panduan Kalibrasi untuk Peralatan Keselamatan Alcolock V3. Alcolock V3 8 halaman. Alcohol gas standard

Alcolock V3 Panduan Kalibrasi
CAUTION!
• It is recommended calibration be done indoors, in a service facility, by maintenance code or
using an ACS download station
• This calibration guide is intended for use by trained technicians only
COMPRESSED GAS SAFETY
• Damaged or broken valves can turn a canister into an unguided missile. Attach the valve in a
safe location
• Calibration should be performed indoors, in a service facility, where the gas standard cylinder
can be properly stored
• Transportation and storage of compressed gases in vehicles is dangerous and should be
avoided
• Examine the canister and valve for any damage, and the expiration date on the label
• Observe all cautions and safety warnings found on the canister
• Never remove or alter canister labels
• Never modify the delivery tube in any way
• Always remove the valve and install the protective cap on cylinders when not in use
• Store cylinders in a cool, well ventilated area away from sources of heat
EQUIPMENT
• 130 PPM (0.05 g/dL) alcohol gas standard, ACS #95-000425
• Regulator valve assembly (6 L/min flow rate): ACS #94-000225 (trigger);
ACS #94-000226 (push button)
• V3 download station, ACS #79-006111

CYLINDER SETUP

Remove the protective cap and screw the regulator valve to the canister. Attach the rigid delivery
tube. Continue to the next section.
CALIBRATION MANUAL
1