Operating Instruction (Without connecting to Smartphone App)
1. Deimos is using a 1800mAh NI-MH rechargeable battery, charge the battery by a micro USB cable, the
red light inside the chamber will turn on when charging.
2. Press and hold the On/Off button for 2 seconds to switch on the device.
3. Main screen:
Next Calibration
4. It is normal for a device has a record around 10 tests, as the device needs to be tested after
calibration.
5. Press the On/Off button to enter the test mode:
6. Insert a mouthpiece, take a deep breath and blow CONTINUOUSLY until a click sound is heard. A
beeping tone will be heard when blowing. Test result will display on the LCD.
7. To make a passive manual test, press the button when the word "Ready" is displayed on the LCD or
during the blow.
Reminder: Passive test cannot be performed when test is done through the App to prevent cheating.
8. Press the button once to perform another test.
Reminder: To perform a test immediately after a POSITIVE test, change the mouthpiece, take an
interval of at least 2 minutes before the next test, or make a passive test with clear air before the next
test to prevent residues.
9. A sampling cup can only be used for screening test, change the test mode to "Screening Mode" in
setting if needed.
10. Press the Left button to return to the main screen.
11. Press and hold the On/Off button for 2 seconds to switch off the device. The device will also be
automatically switched off after 3 minutes.
Present Time
Date
Present Date
YYYY/MM/DD
Sensor
Temperature
Battery Level
No. of Test from
Last Calibration