D Y N A - F O R M M E R C U R Y A D V A N C E
Release the magnet by firstly unscrewing 4
5
(7mm ny-loc) and remove the compressor back
plate and seal.
To remove the shuttle, loosen the compressor screws by
6
approximately 25mm to release the compressor lid and
seal, followed by 8 screws either side of the compressor
casing.
Remove both valve bodies, unscrew and remove
7
the diaphragms.
Replace the diaphragms, ensuring the location nipples
8
are seated correctly in the housing. Confirm the
diaphragm runs smooth around the housing.
The compressor can now be assembled.
9
Attention: ensure no leads are pinched
and seal is intact.
1 4
D I R E C T H E A L T H C A R E G R O U P . C O M
5
6
7
8
S E R V I C E M A N U A L
10
Reattach the compressor to the
10
attachment plate, using 6 new
suspensions.
9.9.2 Rotor Valve Maintenance
The Control Unit has a light that when illuminated indicates a
Service is required. Open the housing by remove the 4 screws from
the bottom part.
Grease the rotor valve: Press down the cap of the rotor valve
release the lock pin. Lift of the cap and put a thin layer of grease
(Silicone grease dow Corning Molykote PG21, 613000103) on the
bottom side of the cap. Assemble back again.
9.9.3 Fuse
Check the fuse: Check the fuse and if required add a spare in the
holder for the fuse in the fuse holder.