Tel: +44 (0)1525 373178/853800, Lo-call: 0845 3108053, Fax: +44 (0)1525 - 851319
Use the table below to input the temperature range limits:
8. Change the value of "Lo.L.1" to the lower limit of the sensor's temperature range, in ºC
9. Change the value of "uP.L.1" to the upper limit of the sensor's temperature range, in ºC
Sensor
PCxxxLT-0
PCxxxMT-0
PCxxxHT-0
PUA8 (all models)
PUA2-151-LT
PUA2-151-PT
PUA2-xxx-MT
PUA2-xxx-HT
PUA5-xxx-GLT
PUA5-xxx-GHT
* The temperature range limits of all PyroUSB sensors may be changed using Calex or
third party software. Please ensure Lo.L.1 and uP.L.1 match the limits you have set using
the software.
10. Press UP and DOWN simultaneously to exit configuration mode. The display now shows
the measured temperature.
Registered Office: 20 Bridge Street, Leighton Buzzard, Beds, LU7 1AL
Registered Number: 3737294 England
PO Box 2, Leighton Buzzard
Bedfordshire, England LU7 4AZ
E-mail: [email protected], Online: www.calex.co.uk
Lo.L.1
-20
0
0
-40*
45*
100*
250*
450*
50*
200*
uP.L.1
100
250
500
1000*
300*
400*
1000*
2000*
1000*
1650*
DRR245 - 4-20 mA Connection Instructions.doc-15:09-29/09/17