Juga untuk Xerox 6100BD - Phaser Color Laser Printer: Panduan Instalasi (34 halaman), Spesifikasi (2 halaman), Panduan Pengguna (32 halaman), Brosur & Spesifikasi (4 halaman), Panduan Pengguna (28 halaman), Instal Manual (2 halaman), Manual Keamanan (38 halaman), Instal Manual (2 halaman), Catatan Rilis (2 halaman), Panduan Memulai (37 halaman), Manual (34 halaman), Manual (34 halaman), Panduan Pengguna (32 halaman), Manual (11 halaman), Panduan Instalasi dan Pengoperasian (8 halaman)

Xerox 6100BD - Phaser Color Laser Printer Panduan Memulai

Reset the Xrite Spectrophotometer

24
Changing the Retain PDL Setting
Perform the following steps to modify the number of PDLs that can
be retained on the DocuSP.
1. On the DocuSP interface, select Setup and System
Preferences. Select the Job Policy tab.
2. Ensure that the setting for Retain PDL files of last 50 jobs is set
as enabled.
3. Select [OK]. Do not restart the system if you are prompted to do
so.
4. In the background, right-click to open the menu. Open a
terminal window.
5. Login as root user.
6. Type: /opt/XRXnps/bin/setretainPDL # and press <Enter>,
where # is a number between 1 and 100. A value of 0
maintains the current setting.
7. Close the terminal window. Restart the DocuSP software.
Perform the following to reset the spectrophotometer:
Ensure that the calibration strip that matches the serial number on
the spectrophotometer is available. If not, order a new one before
continuing.
1. On the DocuSP controller background menu, open a terminal
window.
2. At the % prompt, type:
cd /opt/XRXnps/XRXcalLib/bin
3. From the % prompt, type:
. /resetDTP
This will restore the spectrophotometer to the factory default
settings.
After the reset is complete, follow the standard procedure to
calibrate the spectrophotometer using the Xrite calibration strip.