4. Low-Pressure Indicator
When the mattress is initializing, the Low Pressure LED will illuminate until the appropriate
pressure is reached (according to the pressure adjustment knob); this is normal. Otherwise, the
Low Pressure LED is a warning, indicating the pressure in the mattress is unusually low. Check
to ensure all connections are secured and correctly installed as per the installation instructions.
NOTE: If the pressure level is consistently low, check for any leakage (tubes or air hoses). If
necessary, replace any damaged tubes or hoses or contact your account manager.
5. Alternate/Static Switch
The ALTERNATE/STATIC switch selects between Alternate Pressure mode and Static Pressure
mode. With Alternate Pressure mode, alternating air cells are partially deflated and inflated,
avoiding prolonged pressure on any single point beneath the patient or resident; this is to help
prevent pressure ulcers. With Static Pressure mode, all of the air cells are equally inflated.
6. Alarm Reset
The audible/visible alarm turns on when the pressure is low. To mute the audible alarm, press the
reset button. The visible alarm indicator will flash. Re-press the reset button to reactivate the alarm.
Emergency CPR Operations
CPR must be performed on a firm surface. Therefore, if an emergency CPR situation occurs
with the resident on the mattress, the mattress must be quickly deflated. This is done by quickly
pulling on the CPR tag located at the head of the mattress on the resident's right side. The Quick
Connector found from the pump unit can also be disconnected to hasten deflation.
C
P
R
C
P
R