Nikon NR-5500 Manual Perbaikan - Halaman 19

Jelajahi secara online atau unduh pdf Manual Perbaikan untuk Alat Ukur Nikon NR-5500. Nikon NR-5500 25 halaman. Auto refractometer

Nikon NR-5500 Manual Perbaikan
TEST PROGRAM
© TEST -10: Making functional checks on the touch panel keys
— Press this
key to quit
TEST-10.
• This tests permits the user to check the touch panel keys for normal operation.
♦ Checking the key function
Press the key to be checked. The key is normal if the corresponding display
switches from OFF to ON. The END key is normal when the NR - 5500 quits
TEST -10 normally.
* TEST - 0B may also be executed in conjunction with this test to check the
position of the key display frame.
© TEST -11: Adjusting the cylindrical axis offset
• Model eye alignment adjustment
Place J38056 astigmatism model eye tool on the chin rest.
Focusing
: Focus the blinking dot of the M LED by sliding the Main body
forwards or backwards.
Positioning: Move the Main body by sliding and elevating until the blinking dot of
the M LED is positioned at the center of the alignment mark.
* Perform focusing and position alignment simultaneously. Before installing
the astigmatism model eye tool, clean it with cleaner liquid.
♦ Measuring the cylindrical axis
Press the measurement button to start measurement. The measured cylindrical
axis is displayed in the " AXIS=" field.