• Personal protective equipment (PPE), such as lab coats, disposable gloves and eye
protection, should be used.
• Turn off the device and unplug the power supply plug at the mains.
• Using 75% alcohol disposable wipes, clean the external surface of the apparatus.
• Wait for the instrument to fully dry before reconnecting the power supply.

4.3 Routine maintenance

• Only use a cloth with water, mild soap or a non-abrasive cleaning liquid (pH 6 – 8) to
clean the instrument. Using abrasive reagents may damage the instrument.
• If any reagents or any corrosive substances are spilled on the instrument, immediately
wash with a wet cloth and water.
• The exterior of the equipment should be cleaned regularly and at least on a monthly
basis.
• Good operating practice may extend the life of the product.

4.4 Troubleshooting

The table below lists some common problems
encountered when using the instrument. If the problems are
still not resolved after following the suggested solutions,
please contact a service engineer.
Issue
The device is only
partially working.
Device will not start.
Device is running
continuously
without touch start.
Table 1. Troubleshooting table
www.abbexa.com · Email: [email protected] · Phone: +44 (0) 1223 755950 · Fax: +44 (0) 1223 755951
Possible causes
Small screw in the
bottom may be
loose.
Small screw in the
bottom may be
loose.
Small screw in the
bottom may be
too tight.
Abbexa Ltd., Cambridge Science Park, Cambridge, CB4 0EY, UK
Registered in England: 8475531 · VAT Reg. No: GB 167 9240 79
Solution
Tighten by ½ or 1 turns.
Tighten by 1-2 turns.
Loosen by 1-2 turns
5