Abbott Heartmate 3 Panduan Memulai Cepat - Halaman 2

Jelajahi secara online atau unduh pdf Panduan Memulai Cepat untuk Peralatan Medis Abbott Heartmate 3. Abbott Heartmate 3 4 halaman. Left ventricular assist system

Abbott Heartmate 3 Panduan Memulai Cepat
GREEN
HeartMate 3™ Left Ventricular Assist System
Changing Batteries
WARNING: At least one controller power cable must be connected to a power source
AT ALL TIMES. Do not remove both batteries at the same time or the pump will stop.
l
Obtain two charged batteries from patient's
accessory bag or battery charger. The charge level
of each battery can be assessed by pressing the
button on the battery. Fully charged batteries will
display 5 lights. (Figures 1 and 2)
l
Check the power level on the batteries, replace
the battery with the fewest lights first. Remove only
ONE battery from the clip by pressing the release
button on the clip to unlock the battery. (Figure 3)
Controller will start beeping and flashing yellow
l
symbols and will read CONNECT POWER on the
front screen.
l
Insert a new, fully charged battery into the empty
battery clip by aligning the
battery and clip (Figure 4). The battery will click into
the clip. Gently tug on battery to ensure connection.
If the battery is properly secured, the beeping and
yellow flashing will stop.
Repeat previous steps with the second battery and
l
battery clip.
10
GREEN
System Controller
RED
arrows on the
GREEN
Figure 1
Figure 3
Figure 4
This guide does not supersede manufacturer instructions.
GREEN
Figure 2