3Gen DermLite II PRO HD Petunjuk Manual - Halaman 2
Jelajahi secara online atau unduh pdf Petunjuk Manual untuk Peralatan Medis 3Gen DermLite II PRO HD. 3Gen DermLite II PRO HD 13 halaman.
English
Instructions
CAUTION: Prior to using the DermLite, read these instructions and
keep them in a safe place for future reference.
CAUTION: United States Federal Law restricts this device to sale
by or on the order of a physician.
CAUTION: The DermLite is designed for external examination only.
Introduction
The DermLite II PRO HR is a pocket epiluminescence microscopy
(PELM) device designed to view skin lesions with high magnification
and clarity.
A 25mm 10x lens combined with a unique button-activated light
intensity settings provides for ideal skin microscopy. A high quality
lens with superb color correction and reduced image distortion is
used to image the skin. To facilitate the use of immersion fluids,
and allow for fixed digital imaging, the unit is equipped with a
retractable faceplate spacer.
Directions for Use
CAUTION: Do not look directly into the bright LED light. Patients
must close their eyes during facial examinations.
IMPORTANT: Before initial use, please charge the internal lithium
ion battery for a minimum of 4 hours using the charger provided.
Hold the DermLite II so that the LEDs are aimed in the direction
of the lesion to be examined. Place the device approximately 1"
(~25mm) from the skin. Push the power button (A) at the left side of
the unit for 2-3 seconds to activate the LEDs.
Look through the lens with your eye placed 1"( ~25mm) to 4"
(~100mm) above the DermLite II. Move the device closer or further
away from the lesion to obtain the desired image focus. For oil
immersion imaging, or for camera use, rotate the spacer dial (B) so
that the white marker is at the 12 o'clock position. For fine focusing
within a range of +/- 1mm, simply rotate the dial to the left or right
between the 11 and 1 o'clock markers (C).
CAUTION: If the DermLite II PRO HR is used for oil immersion
imaging, place the oil only on the skin and then place the extended
faceplate spacer on top of the oil. Do not allow oil to contact the
USB port, DC charging port, or battery compartment
To attach the DermLite II PRO HR to a camera, the device is
equipped with a 28mm thread (D). A simple ring adapter (such as a
28-37mm stepping ring) may be required to adapt the device to your
camera of choice. Please contact 3Gen or a photographic retailer
for more information.
Removable Faceplate (E)
The DermLite II PRO HR is equipped with a removable faceplate. To
remove this piece, rotate the spacer dial (B) to extend the spacer,
and twist off the small ring in the very front of the spacer.
Light Intensity
A push on the button on the right side labeled "MODE" (F) toggles
between two light intensity settings. The first mode activates 16
LEDs, and the second mode gives you twice as much light with the
activation of 32 LEDs.
Rechargeable Battery
After a minimum 4-hour initial charge, the lithium-ion battery
provides up to 2 hours of continuous operation, or enough power for
imaging approximately 600 lesions at 30 seconds per examination.
Once the battery is depleted, the unit will automatically shut off, and
will need to be recharged. Plug the charging cord into the round DC
charging port (G), and the red charging indicator (H) will illuminate.
Once the battery is fully charged, the green indicator will light up (I).
Optionally, you may use a USB cable to charge the unit via the USB
port (J).
CAUTION: When charging with a USB cable, the DermLite II PRO
HR must be connected only to the USB port of a computer that
complies with the IEC / EN 60950 standard to ensure continued
compliance of the product with the European Medical Device Direc-
tive. Refer to the operator's manual provided with your computer to
ensure that the computer complies with the above standard.
After approximately 1,000 charges, the battery's capacity will dimin-
ish and the battery may need to be replaced. To remove the battery
door (K), push down on the two ovals (L) with your thumbs and slide
the battery door back. Disconnect the small white battery plug, and
remove the battery. Insert a new battery and reconnect the battery
plug. Replace the battery door securely on the device.
The DermLite II PRO HR uses a special lithium-ion battery, which
can only be purchased from 3Gen or from an authorized 3Gen
dealer.
CAUTION: Do not under any circumstances use another battery
other than the one designed for this unit. Doing so could result
in damage to the unit. Also, do not under any circumstances use
another charger other than the one supplied with the unit.
Troubleshooting
Please check our website (www.DermLite.com) for the most cur-
rent troubleshooting information.
If you need to return the DermLite II PRO HR to 3Gen for servicing,
please call 3Gen prior to returning the device, and a unique Return
Merchandise Number (RMA) will be issued to you.
No Light
Fully recharge the battery and check again. If the condition persists,
send the unit to 3Gen for evaluation and/or repair.
Weak Lighting Recharge the battery.
LED Fails
The LEDs used in the DermLite II PRO HR are designed to last over
100,000 hours. If any of the 32 LEDs fail, contact 3Gen for repair.
Mechanical Malfunction
Please contact 3Gen to arrange for service.
Care and Maintenance
The DermLite II PRO HR is designed for trouble-free operation.
There are no adjustments to be made nor is any special servicing
required. Never attempt to open the device for any reason.
Cleaning and Sterilization
The DermLite II PRO HR body may be cleaned or disinfected by
wiping it with Isopropyl alcohol (70% vol.). Do not use alcohol or
disinfectants in the optical areas of the unit. Do not use abrasive
material on any part of the equipment or immerse the device in
liquid. Do not autoclave.
Cleaning the Lens and the Polarizers
The lens and the polarizing filters should be treated as high quality
photographic equipment and should be cleaned with standard lens
cleaning equipment and protected from harmful chemicals.
Specifications
The patent-pending DermLite II PRO HR features an innovative opti-
cal microscopy system designed for superb clarity and outstanding
vision.
DermLite II PRO HR components include:
• Thirty-two 3mm white LEDs
• 25mm lens with 10x magnification
• Two light intensity settings
• Retractable Faceplate Spacer with 10mm Scale
• Lithium-Ion Battery
Warranty
5 years for parts and labor.
Serial number and company information are located inside the bat-
tery compartment.