1-CUBE SU 6.2. Panduan Pengguna - Halaman 13

Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Termostat 1-CUBE SU 6.2.. 1-CUBE SU 6.2. 13 halaman. Water thermostat

6. M
ONITORING
During all the methods Heating Immersion Circulator the records of the temperature curves in PC are possible. Switch
off the device and connect the PC com port with the RS232 com port with the help of supplied communicating cable.
Switch on the device and start any method. Use any program which makes possible the RS232 communication to
record the temperatures in PC. Program Hyperterminal (which is part of WINDOWS) or program Serial Watcher (which
is part of Monitoring Mash Bath supply) can be used. Note. It is necessary to adjust the program for ASCII signs
reception.
7. M
AINTENANCE
Throughout the operation the scale can be sedimented on the heating elements. It is important to remove these
sediments. Pour the water with either the acetic acid or citric acid into the bath and heat up and hold the
temperature of 65 degrees Celsius for one hour in the bath. Then disconnect feed el.cord out of the socket, suck the
sediments with the help of a hose and clean the vessel.
8. I
MPORTANT WARNINGS
The Cooling Bath must be always located on the nibs and mustn't be tilted or turned upside down
in any case!!!!
If the device is transported at temperature below freezing point, it is necessary to leave the instrument for about 2
hours at room temperature before you start to work with it.
The fuses are located on the back panel of the Water Bath. It is forbidden to replace the fuse by fuse with different
value than the value indicated on the equipment.
Service:
For service contact the manufacturer:
NOTICE: It is forbidden for anyone except for the manufacturer or authorized company to repair the apparatus.
At the end of the apparatus life we recommend that it is according to corresponding waste categorization.
Take the metallic parts made of iron, brass, plastic elements and packing material to the separated collection as a
secondary raw material. You can also order the apparatus liquidation at the above mentioned company that ensures
the liquidation of the waste according to Waste Act.
RS232
1-CUBE, Hamry 3567, 580 01 Havl.Brod, Czech Rep.
tel. 00 420- 569 433 620
fax .00 420-569 422 144
13