Etac Molift Air Series Panduan Pengguna - Halaman 8
Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Bantuan Mobilitas Etac Molift Air Series. Etac Molift Air Series 15 halaman.
Molift Air
General Safety Precautions
Only use accessories and slings that are adjusted
to fit the user, type of disability, size, weight and
type of transfer.
Working pause ratio/Duty Cycle.
Molift Air should not be run constantly for more
than 2 minutes (with maximum load), and rest
for minimum 18 minutes. Duty Cycle 10%.
(Intermittence according to standard ISO-EN
10535)
Before use / Daily check
Inspection to be performed daily or before use:
Inspect lifter and rail system has no visible
damage, defects or deformations
Make sure suspension connection and all
detachable parts are properly connected and
secured
Test emergency stop button and emergency
lowering
Make sure battery indicator or service light is
not illuminated
Hand control
The Hand control has 2 buttons for lifting and
lowering, or 4 buttons if the hoist is equipped
with propulsion. The Hand control has an indica-
tor light that will illuminate when battery level is
low and the lifter requires charging
How to use Molift Air
Safe Working Load
Make sure lifting strap does not have visible
damage or frays.
Test run lifters operation functions and make
sure lifter does not make any abnormal
sounds
If there are any faults or defects, the lift needs to
be taken out of operation and marked "out of
order"
7
If maximum load (SWL) differs
between hoist suspension and body
support unit, then the lowest maxi-
mum load shall always be used
Molift lifters shall only be used to lift
persons. Never use the lifter to lift or
move objects of any kind.
Do not pull Hand control to move
the hoist along the rail