Chapter 1 – Overview
The DOD Technologies ChemLogic 8 (CL8) Pyrolyzer option uses two four-point Pyrolyzers enclosed in a
separate housing the same size & shape as the CL8. The Pyrolzyer has it's own power cord, on/off switch,
and exhaust fan as shown in figure 1.
Tubing and electrical
connectors from CL8.
Points 1-4 Pyrolyzer
WARNING:
To provide adequate cooling of the unit provide a minimum of 6" of
clearance on both sides of the Pyrolyzer unit for proper air circulation. Do not
block the exhaust fan.
1.1 Pyrolyzer Installation Diagram
CL8 Operating Voltage: 110/115VAC (+/-10) 50/60 Hz <1.0 Amps;
230VAC (+/-10) 50/60 Hz <0.5 Amps
Pyrolyzer Operating Voltage: 110/115VAC (+/-10) 50/60 Hz <2.0 Amps;
230VAC (+/-10) 50/60 Hz <1.3 Amps
Sample Tubing Points 1-8 FEP Teflon: ¼" OD x 3/16" ID x 300' (91 mtrs)
Exhaust Tubing Polyethylene: 3/8" OD x 1/4" ID x 25' (7.5 mtrs)
Please allow 4' minimum spacing (left and right side) to allow door swing and power plug clearance
DC-ITD-CL4-CL8MANPYRO_A
REV 8/25/2022
Points 5-8 Pyrolyzer
Figure 1
3
Exhaust fan
On / Off Switch
Power Cord
Internal Power Supply
CL4 & CL8 Manual