2. Replacement of Needle
If needle is no longer sharp or in straight condition, it should be replaced.
Use a small screwdriver 3mm to remove the lower small screw, see figure 7.
Replace the needle with a new according to the specified part in the spare part list below.
When mounting the new needle, use some silicone grease on screw and needle to
facilitate future replacement. Do not over-tighten the screw.
Replace the Spring Stop by removing this screw
Replace the Leaf Spring by removing this screw
Replace the Needle by removing this screw
Figure 7: Replacement of de-capper head spare parts
3. Replacement of Spring stop
If the spring stop is damaged and no longer could be adjusted in line according to the
figure 8, it should be replaced. Use a screwdriver 5mm to remove the upper screw, se
figure 7. Replace the Spring Stop with a new according to the specified part in the spare
part list below. Check correct mounting with figure 8.
4. Replacement of Leaf spring
If the leaf spring is no longer holding the puller in a tight position when head is removed, it
should be replaced. Use a small screwdriver 3 mm to remove the middle screw, see figure
7. Replace the Leaf Spring with a new according to the specified part in the spare part list
below.