Aseptico AEU-40 Endopex V Panduan Instruksi Pengoperasian dan Pemeliharaan - Halaman 4
Jelajahi secara online atau unduh pdf Panduan Instruksi Pengoperasian dan Pemeliharaan untuk Peralatan Medis Aseptico AEU-40 Endopex V. Aseptico AEU-40 Endopex V 12 halaman. Vector based root apex locator

OPERATION FUNCTIONS:
1. Power Switch – Controls power on/off to the unit. When
turned on, self-test will initiate and should illuminate a green
power L.E.D. within a few seconds to indicate a properly
functioning unit.
a. A flashing yellow L.E.D. indicates a low battery requiring
replacement.
b. The unit will automatically turn off if it fails the battery of
power on tests. Low batteries will cause self test failures.
c. If a subsequent attempt to power on succeeds, the unit
will function properly, but should still be returned for serv-
ice if it continues to automatically turn off during self test
with fresh batteries. Low batteries will cause self test fail-
ures.
2. Display Switch – Allows the selection of one of three dis-
play modes. The following modes are indicated by the illu-
mination of a corresponding L.E.D. color below the Display
Switch;
a. Single Bar March (Green) – One bar is illuminated to
indicate proximity to the apical foramen. This mode is
most conservative of battery life.
b. Logarithmic March (Yellow) – Further away from the api-
cal foramen, 3 bars will light, mid range proximities will
produce 2 illuminated bars, near proximities, apex indica-
tion and perforation indications will illuminate only one
bar or light. This mode is also conservative of battery life.
c. Persistent On March (Red) – Each illuminated bar
remains on during progression toward the apical fora-
men. This mode is least conservative of battery life, but
certainly gives the most visible indications.
3. Sound Switch – Controls sound indication on/off. The state
of the sound indication is indication by the on/off illumination
of an L.E.D. beneath the Sound Switch. When active the
following audio indications will be provided;
a. Distant proximity – silence.
b. Near proximity – 1/24th pulse chirp.
c. Impending proximity (one light bar preceding the Apex
indicating 0.5mm) – 1/3rd pulse beep.
d. Apex – 1/2 pulse beep.
e. Perforation – solid tone alarm.
2.