TROUBLESHOOTING:
Problem:
Console does not light when
turned on:
Console lights when turned on,
but handpiece does not turn:
No Water flow from pump to
handpiece:
SYMBOL DEFINITIONS:
Dangerous
Type B
Voltage
Equipment
The UL 2601 Standard Duty Cycle for Intermittent Use is 1 min. on / 5 min. off.
Correction:
• Check console to power connection.
• Check that Voltage Selector is set for the proper
voltage.
• Check fuse. If blown, replace with 1A/250V
slo-blo fuse for 110 volts, & 0.5A/250V slo-blo
fuse for 220 volts.
• Check motor plug connection.
• Check foot switch connection.
• Depress foot switch.
• Increase RPM.
• Check that bur is properly seated in the
handpiece and the collet is closed.
• Check that pump is on and flow level is sufficient.
• Check that water container seal is completely
punctured.
• Make sure the irrigation tubing is properly installed
in pump and is in the correct direction.
Attention - Consult
Accompanying
Documents
7