Aseptico Sealant Express ADU-08SXE Petunjuk Pengoperasian - Halaman 4

Jelajahi secara online atau unduh pdf Petunjuk Pengoperasian untuk Penyedot Debu Aseptico Sealant Express ADU-08SXE. Aseptico Sealant Express ADU-08SXE 4 halaman.

Aseptico Sealant Express ADU-08SXE Petunjuk Pengoperasian
MAINTENANCE:
Air/Water filter- The ADU-08SXE unit contains a
built-in air line filter. It removes unwanted moisture,
oil, dirt, chips, and other contaminates that may enter
the air supply line. To ensure clean dry air to your
unit, the filter bowl must be drained regularly. If a
significant amount of water can be seen through the
sight glass on the moisture collection bowl, the filter
should be drained. To drain, extend the drain line
outside the unit and unscrew the drain valve. Allow
all moisture to drain out, then retighten drain valve.
Surface Disinfecting- Any external surfaces of the ADU-08SXE unit that
are contacted during use should be carefully wiped down with a disinfectant
between patients. The entire unit should be thoroughly cleaned at least once
a day.
After each patient, the air vacuum system should be thoroughly rinsed by
evacuating clean water through the hose end followed by air to clear all water
from the tubes. The entire system should be cleaned and disinfected each
time it is emptied.
Syringe tips should be changed between patients, and used tips thoroughly
cleaned and autoclaved before reuse.
Water Lines- Disinfect the water lines weekly. Prepare a 1:10 bleach
solution (1 part household bleach to 10 parts water). Remove water reservoir
and discard residual water. Replace empty water supply tank and air purge
all waterlines. Fill water supply tank with bleach solution. Run bleach solution
through all lines. Allow bleach solution to stand in lines for 10 minutes.
Remove water supply tank and discard bleach. Flush water supply tank and
all lines thoroughly with clean water. Air purge and leave lines dry until next
clinical use.
Warranty
Aseptico warrants its products against defects in material or workmanship for a period of two (2)
years, from date of original invoice. Some handpieces are warranted for one year under the
same conditions. Other handpieces and expendable components, such as air turbines and light
bulbs, are covered by shorter warranty periods, or have no warranty. Aseptico's sole obligation
under product warranty is (at its sole option and discretion) to repair or replace any defective
component or product in part or whole. Aseptico shall be the sole arbiter of such action.
In the event of alleged defect under warranty, the purchaser is to notify Aseptico's Customer
Service Department promptly. Customer Service will provide instructions, usually directing
that the product be returned for service. Shipment to Aseptico and the cost thereof is always
the responsibility of the purchaser.
Accidental misuse, inappropriate installation, or failure to perform directed maintenance voids
the warranty. Deliberately defacing, modifying, or removing the serial number voids the warranty.
Aseptico does not assume, under this warranty, any risks or liabilities arising from the clinical
use of its products, whether or not such use involves coincidental utilization of products
manufactured by others.
EC
Advena Ltd.
0086
Pure Offices, Plato Close, Warwick CV34 6WE UK
REP
FILTER
SIGHT
GLASS
UNSCREW
HERE
AIR SUPPLY LINE
DRAIN
TO COMPRESSOR
MOISTURE