EG&G ORTEC 535 Panduan Pengoperasian dan Servis - Halaman 9

Jelajahi secara online atau unduh pdf Panduan Pengoperasian dan Servis untuk Penguat EG&G ORTEC 535. EG&G ORTEC 535 10 halaman. Quad fast amplifier

BIN/MODULE CONNECTOR PIN ASSIGNMENTS
FOR AEC STANDARD NUCLEAR INSTRUMENT
MODULES PER TID-20893 (Rev 4)
(adopted by DOE)
Pin
Function
Pin
Function
1
+3 volts
23
Reserved
2
-3 volts
24
Reserved
3
Spare Bus
25
Reserved
4
Reserved Bus
26
Spare
5
Coaxial
27
Spare
6
Coaxial
•28
+24 volts
7
Coaxial
•29
-24 volts
8
200 volts dc
30
Spare Bus
9
Spare
31
Spare
•10
+6 volts
32
Spare
•11
-6 volts
•33
117 volts ac (Hot)
12
Reserved Bus
•34
Power Return Ground
13
Spare
••35
Reset (Sealer)
14
Spare
••36
Gate
15
Reserved
"37
Reset (Auxiliary)
•16
,+12 volts
38
Coaxial
•17
-12 volts
39
Coaxial
18
Spare Bus
40
Coaxial
19
Reserved Bus
•41
117 volts ac (Neut.)
20
Spare
•42
High Quality Ground
21
Spare
G
Ground Guide Pin
22
Reserved
Pins marked (*) are installed and wired in EG&G ORTEC's 4001A. 40018.
4001C. 401A. and 4018 Modular System 8ins.
Pins marked (*) and (") are installed in EG&G ORTEC-HEP M250/N
and M350/N NIM8INS.