CytoTherm CT-4T Panduan Pemilik - Halaman 2

Jelajahi secara online atau unduh pdf Panduan Pemilik untuk Peralatan Laboratorium CytoTherm CT-4T. CytoTherm CT-4T 10 halaman.

CytoTherm CT-4T Panduan Pemilik

DESCRIPTION

The CytoTherm-4T (CT-4T) is a laboratory instrument that consists of a recirculating water temperature bath
that uses a rocking action to generate bursts of flow to massage the plasma. This massaging action speeds up the
thawing time.
The CT-4T comes with one CORRAL that will thaw 1 to 6 plasma bags at a time. A second CORRAL will let you
thaw 6 more plasma bags for a total of up to 12 plasma bags at a time. The operator never comes in contact with
the water in the bath when handling the plasma bags. The CORRAL keeps the entry ports of flat- frozen units out
of the water so an overwrap bag is not necessary. Folded frozen plasma units require an overwrap bag to protect
the entry ports. Both type of freeze methods can be thawed together.
A large digital display indicates the temperature and counts down the thawing time. The CT-4T comes with a
clear evaporation lid.

SETTING-UP

The unit comes packed in 1 box. Make sure you have received the following:
..... Control Panel
..... Corral with 6 Bag Clamps
..... Drain Tubing
..... Power Cord
..... Overwrap bags
......Algaecide. USA only
Instructional video can be found at www.cytotherm.com
Screw the ARMS to the side of the CONTROL SECTION. The notched side of the ARMS face up.
Place the BATH in front of the CONTROL SECTION so that the chassis on the BATH is next to the indented
portion of the CONTROL SECTION.
Connect the umbilical cord of the CONTROL SECTION to the plug receptacle on the back of the BATH section.
WHEN CONNECTING OR DISCONNECTING THIS CONNECTION * *
Lift the BATH and position it so that the pivots on the side of the BATH slip into the notches on the ARMS.
Make sure the power switch is OFF. Connect the power cord to a properly grounded electrical outlet. Mount the
CORRAL inside the BATH. Push The SLIDERS on both sides of the BATH towards the front of the BATH. In
this SLIDER position the CORRAL can be placed on the BACK SHELF. Mount the first CORRAL towards the
back of the BATH. The cutout of the CORRAL should be on the right side. Push the SLIDER to its middle
position, this will hold the CORRAL in place.
Second CORRAL (optional accessory). If you are mounting a second CORRAL to thaw up to 12 plasma bags at a
time, push the SLIDERS towards the back of the unit and mount the second CORRAL on the front SHELF. The
second CORRAL mounts like the first CORRAL. Move the SLIDER to the center to hold both CORRALs.
..... Bath
..... 2 Arms
..... Evaporation Lid
......Instruction Book
..... Drain tubing with male drain fitting
* * THE POWER SWITCH MUST BE OFF
2