Calibration

1. Flow 100% Pure Nitrogen (N2) (0% CO2) into CO2 Sensor Entry (Item A in
Layout). Wait 3-5 minutes before starting calibration, then execute the "ZEro"
calibration.
(Continue to flow calibration gas throughout entire process)
2. Press Mode until CALI is flashing in small print under the CO2 reading.
3. Press Enter to view the calibration settings.
4. "ZEro" and "CO2" will be flashing.
5. Press Enter again to begin the calibration. "CALIBRATING" will begin flashing.
(Continue to flow N2 throughout the entire calibration)
6. After approximately 3 minutes, "PASS" or "FAIL" will be displayed.
7. If "PASS" press Enter to save.
8. If "FAIL" repeat the process.
9. If "FAIL" more than twice, contact CO2Meter technical support.