Sfoglia online o scarica il pdf Manuale d'uso per Strumenti analitici Aktakom AVS-1050. Aktakom AVS-1050 10. Video borescope
Image signal system
•
Signal path: there are two accesses of the image signal entering, one way
is for the image signal of inside camera and the other is for the external
AV signal.
•
Signal Switching: The meter's signal access is default of inside camera,
on the TFT-LCD it display the illumination level indication, so when an
external signal comes, the unit will automatically to switch to external
access within 5 seconds, illumination level indication will disappear.
when the external signal is cancelled, the system automatically switch
to the internal access road within 2 seconds.
•
Signal format conversion: Internal signal format defaults to PAL, without
doing the signal format detection, external signal format can be converted
to NTSC / PAL automatically.
•
Auto Power off if no signal: When there isn't any internal signals or external
signals to access, TFT-LCD displays NO SIGNAL, LED flashes green, and
the meter is in standby, so if there is signal now, it will wake up the meter
and make it enter into working mode. But if there isn't signal within 1.5
seconds, the meter will automatically power off.