Barnstead Thermolyne Corporation DB5910E Manuale operativo ed elenco parti - Pagina 5

Sfoglia online o scarica il pdf Manuale operativo ed elenco parti per Accessories Barnstead Thermolyne Corporation DB5910E. Barnstead Thermolyne Corporation DB5910E 16. Dri-bath test tube incubator

Barnstead Thermolyne Corporation DB5910E Manuale operativo ed elenco parti

Electrical Ratings

Model
Volts Amps Watts
DB5915E
120
.38
DB5925E
120
.38
DB5910E
240
.19
DB5920E
240
.19
Factory Calibrated Temp.: 37°C ± .25
Operating Range: 30°C to 60°C
Temperature Uniformity: ±.20°C at 37°C
Temperature Stability: ±.30°C at 37°C
Heat-up time of unloaded dri-bath from a cold start to 37°C: 15 minutes for 10 well units, 20 minutes for
20 well units.
Ambient Operating Range: 17°C - 27°C
Repeatability: ±.25°C

General Specifications

Hz
Wells
45
50/60
10
45
50/60
20
45
50/60
10
45
50/60
20
5