5-2. Installation
Before
installation,
please
WARNING/CAUTION].
There are two ways to install the heat flow sensor:
The installation method will differ by installing permanently or temporarily.
Before installing the heat flow sensor, heat flux direction and polarity of the heat flow sensor output must be
verified (see [5-2. Parts Name and Descriptions]). Polarity of the output voltage when heat flow enters from
the black surface with model number (MF-180, HF-30S, HF-10S) of each heat flow sensor or surface with a
white dot (MF-180M) is indicated
1. Permanent Installation
Embedding the heat flow sensor in a measuring surface
This method of installation uses adhesive; once the heat flow sensor is adhered to the surface, the
NOTE:
sensor cannot be peeled off or reused.
1.
Place the heat flow sensor on the measuring surface so that the heat flows perpendicular to the heat flow
sensor, and attach it with an appropriate adhesive* or similar material used on the subjected surface.
2.
Avoid any air gaps between the heat flow sensor and the subjected surface; embed the sensor using the
same material used on the subjected surface.
3.
To avoid getting effects from the surrounding radiation, paint the surface with same color.
*Adhesive, such as One Component RTV (deoximation type), KE45W by Shin-Etsu Chemical Co.
2. Temporary Installation
Attaching the heat flow sensor on the measuring surface
1.
Place the heat flow sensor on the installing surface and attach it with thermal conductive silicon grease**.
Avoid using such as double-sided tape which would require strong force to remove the heat flow sensor
from the surface.
2.
Avoid any air gaps between the heat flow sensor and the subjected surface.
3.
To avoid getting effects from the surrounding radiation, paint the surface with same color.
**Grease type silicon grease has high thermal conductivity and bonding effect. Optional accessories:
YG6111 (200g、1kg)