dacell SM150 Manuale - Pagina 2

Sfoglia online o scarica il pdf Manuale per Strumenti di misura dacell SM150. dacell SM150 4. Digital indicator
Anche per dacell SM150: Manuale di istruzioni (12 pagine)

3. DISPLAY and KEY operation
1)DISPLAY : 4½ Digit LED
* Key operation start : press to
Key operation exit : press to
* Display : setting mode and operation mode.
1. Setting mode : data confirmation or correction.
2. Operation mode : normal operation state.
3. Press to
key : setting mode.
Press to
key : operation mode.
2) L.E.D Display
LED state
operation
Display operation mode of high relay.
HI
setting
Confirm and adjust high data
operation
Display operation mode of low relay.
LO
setting
Confirm and adjust low data
setting
Confirm and adjust full scale data.
FS
Input the full range or capacity of sensor.
DP
setting
Confirm and adjust decimal point.
key in sensor output data ( mV/V )
CAL
setting
* refer to 6 - 1)
setting
auto calibration execution mode.
FS+
cal data is adjusted automatically.
* refer to 3 - 6)
CAL
setting
Adjust display time and unit of
F S +
measurement data
DP
* refer to 6 - 2)
setting
Software adjustment mode.
DP+
* refer to 6 - 3)
CAL
F S +
operation
Set or reset data lock mode
DP+
* refer to 3 - 4)
CAL
3) KEY operation
Mode
1. HI LO FS DP CAL FS+CAL FS+
select
DP DP+CAL
2. All FND of selected mode will blink.
3. Press key : 10000→1000→100→10→1
Positio
Selected digit position will blink
n
4. DP(decimal point mode) : decimal point
select
position movement key.
10.000 100.00 1000.0 10000
Number
5. Data correction key.
select
6. 5 digit
7. The others 1 2 3 4 5 6 7 8 9
0
Input
8. Input data. After setting up must press
key therefore become input.
9. Sensor : No load state(0V, 4 )
A u t o
10. Display : 0
zero
key.
key.
Description
1, -1, -, non
→ → → → → → → →
Key result
Lock
11. Protect setting data.
mode
12. Reset lock mode to adjust data
4) PROTECTION mode
Use this mode for protection and security of System.
high, low data is possible to adjust irrespectively.
Mode
* Press
and
Unlock
time
FS, DP, CAL LED will light up and reset lock mode
(data correction enable)
* Press
and
time in unlock mode : FS, DP, CAL LED turn off
Lock
(lock mode)
* power off : Set lock mode automatically.
5) AUTO ZERO mode
Initialize sensor and indicator.
Execute this mode at unlock mode.
Execution process (display : operation mode)
(1) Sensor : no load.
(2) Voluntary data display.
(3) Press
key and
Zero display (ex : 0, 0.0, 0.00, 0.000)
(4) This data is memorized to indicator though power turn off.
(5) If operate mode with input signal to remain, this point is
memorized zero point.
6) AUTO CALIBRATION mode
Convenience to use when do not know calibration data of
sensor or displayed the measurement value differently.
EX)
At sensor Supply : 100.00 kg/㎠
Display
Adjustment
(1) Adjust zero point of indicator.
(2) Apply known load to sensor. Large load is high safety.
(3) Press
key : F.S and CAL L.E.D turn on at the same
time.
(4) As example table, 100.05 or 99.05 will blink.
(5) After correction of 100.00 using
press
key.
(6) 100.05 is display correction of 100.00.
(7) At this time, cal data is corrected minutely.
(8) If corrected cal data is over limits of 2-4), correct of DIP
S/W setting and again set the auto calibration.
4. Operating flow chart
Install and connection --Mount indicator to panel and make
connection. (Check AC POWER 110/220V)
Unlock mode -- Press
DATA Input
-- Set data(high, low, F.S, DP, CAL),
refer to 6
DIP S/W setting --sensor output data ---DIP S/W 1, 2, 3, 4
analog output
filter frequency ---- DIP S/W 6, 7, 8
Auto zero
-- Sensor : no load
Press
AUTO CAL
-- Execute auto calibration.
Mode Check
-- Check function(Relay output,analog output,
reset, hold, peak).
Lock mode setting --Press
Description
confirm and setting
key during 5 sec at the same
key during 5 sec at the same
key at the same time.
: 100.05 or 99.05 kg/㎠
: 100.00 kg/㎠
key and
key(data protection unlock)
---- DIP S/W 5
key (auto zero)
key(data protection mode)
key,