General guidelines

To disassemble and assemble the columns follow the instructions below. After reassembling
the column ensure that the column does not leak by performing a hydrostatic leakage test.
When assembling/dissembling attention should be paid to:
1.
Always use the correct size tools. A comprehensive tool kit was provided with the
original column. See Table below.
2.
Ideally components should be at approximately room temperature
(approximately 20° C). If not, take particular care as thermal expansion or contraction may
make components fit more tightly and thus more difficult to disassemble.
3.
Dismantled assemblies should be rested on a clean, dry surface
4.
Pay particular care not to damage the edge of the flowcells
5.
When unpacking columns be careful not to scratch the inside of the column tubes
6.
Be careful not to damage stainless steel surfaces by bumping, knocking or scratching
7.
Do not over-tighten components and use correct torque settings as given in this guide.
8.
Good engineering practices should be followed at all times and operations carried out in
an approved and safe manner.
Tool
Allen Key Set
Allen key Adaptor
Spanner 13mm / 17mm open ended
Spanner 22mm/24mm
Fixed End Tool - Large
Torque Wrench 3/8" Drive
Spirit Level
Micro S/Driver Slotted 1.2mm
Various locations
Flange Fixing Bolts
13mm spanner - column flow tube connection on
adjuster flowcell
17mm Spanner – flow tube on fixed flowcell
Adjuster Feet
Used with torque wrench on fixed end cell fixing nuts
Used to set correct torque on fixed end fixing nuts
Used to level the column
To assist removal of bed support seal assembly
6
Where used