caused by the patient while on the scale. Although this weight may be less
accurate than a Stable Hold weight, the deviation from the patient's actual weight
will be minimal.
The time required for the Average Hold to engage is based upon the Average
Aperture parameter. See the Parameter Setup section for configuration
instructions.
Battery Indicator
The battery indicator displays the remaining battery life available. See the battery
operation section for detailed information. NOTE: Battery Indicator will be disabled
when using an external power supply.
Software Number and Revision display
Pressing and holding the Reweigh button for 3 seconds will perform a display test,
show the Software number "Su181" and the revision "r 3.1" or higher and then
display "StESt" to indicate Switch test mode is active. The Switch test mode stops
after 10 seconds if no switch was pressed. Scale will then run a Display Segment
test. After Display Segment test, the scale will return to the normal weighing
operation.