ASTRO-MED Dash 16u Manuale di avvio rapido - Pagina 11
Sfoglia online o scarica il pdf Manuale di avvio rapido per Registratore vocale ASTRO-MED Dash 16u. ASTRO-MED Dash 16u 13. Field and lab recorder
I. Real-time Recording
DISCUSSION
ACTION
Default the Dash 16u
Create a chart with one signal
Select Chart with Channel 1
Set speed to 5 mm/sec
Stop the chart paper
Set Channel #1 to be +/- 4 Volts FS
Input sine wave into Channel #1
Run the chart
Stop the chart paper
Measure the sine wave signal
using the color display
Remove sine wave signal
Exercise #2: Sine Wave Generator Test
Items needed: Waveform Signal Generator. Signal input leads.
Input a +/- 2 Volt sine wave, 1 kHz into Channel #1. Use a one grid chart.
Set up the channel so that when there is no signal, the waveform trace will be at
the center. When the sine wave signal is applied, the waveform trace will be
shown symmetrically within the grid. The signal will be seen on the color
display as well as on the chart paper. Next, a data capture will be set up to
collect data when the sine wave signal is applied.
[SYS] [SYSTEM SETUP] [SYSTEM DEFAULT] [YES]
[SYS] [AUTO-CHART SETUP] [NEXT >>] To begin auto-chart.
Keep chart format to individual [NEXT >>]. Turn encoder wheel so that
number of channels is 1 [NEXT >>]. Keep chart with channel text [NEXT >>].
Turn encoder wheel to Chart 4 [NEXT >>] [NEXT >>]
[MODE] [RUN CHART 4]
[5] [SEC] Common speeds have dedicated keys on front panel of Dash 16u.
[RUN / HALT]
[AMP 1] One key for each of two amplifiers.
[ADVANCED]
[CHANGE AMP TYPE...] The Dash 16u has four amplifier types.
[SET TO SINGLE-ENDED] Select the single-ended amplifier type.
[EXIT] [SET TO 40 V] Set the attenuator range to 40 Volts.
[EXIT] [SET RANGE] Turn encoder wheel until range is set to 8 Volts.
[SET OFFSET] Turn encoder wheel until offset is set to 0 Volts.
[EXIT]
Use the banana jack input connection located on the side of the Dash 16u.
[RUN / HALT] The signal for Channel #1 should be
[RUN / HALT]
[FREEZE DISPLAY] Toggle the time cursor button until it reads
[TIME CURSOR A]. Turn the encoder wheel to move Cursor A. Observe the
voltage readings shown at the bottom of the display. Experiment with Cursor B
and then Cursors A+B. Amplitude and time measurements between the cursors
are given on the bottom of the display.
When finished, run the display [RUN DISPLAY]
HOW TO
½
of the full scale chart.