COKEVA 11940C Nota operativa - Pagina 10

Sfoglia online o scarica il pdf Nota operativa per Strumenti di misura COKEVA 11940C. COKEVA 11940C 19. Close field probe

 
3‐8.       Power Density and Electric Field Calculations 
 
3‐9.       The Cokeva 11940C measures magnetic field strength. Assuming a 377‐ohm field impedance, an associated 
electric field strength or a wave power density can be calculated from the magnetic field measurement. Refer to the 
following equations: 
E─Field Strength 
 
Power Density 
 
 
 
 
 
 
 
 
EXAMPLE: 
 
  then 
 
E  
 
 
 
3‐10.    Localized Field Source 
 
3‐11.    The probe can also be used as a localized magnetic field source for locating susceptibility problems on an 
electronic circuit. It is a reciprocal device; that is, a voltage fed into the unit will create a magnetic field at the dual‐
loop sensor. For this application, the maximum input power is 0.5W with a characteristic worst‐case VSWR of 3:1. 
Table 2 lists characteristic magnetic field strengths developed at the probe tip using 0 dBm incident power. 
 
Frequency 
(MHz) 
30 
100 
500 
850 
1000 
 
 
 
 
 
(
)
µ
 
dB 
= V
 + AF + K + 51.5 
SA
m
  = V
+ AF + K + 51.5
SA
(
)
mW
dB 
 
 = V
 + AF + K ‐ 104.3 
SA
2
cm
 
where 
 
V
 = voltage in dBµ V measured on analyzer
SA
AF = 11940C Antenna Factor in dB
K = loss or gain in dB between 11940C and spectrum analyzer 
(Positive value if loss, negative value if gain.)
At 100 MHz, measured V
(
)
µV
 
dB
 
= 60.3 + 44 + 51.5 = 155.8 dBµV/m = 61.7 V/m 
m
(
)
mW
 
dB
 
= 60.3 + 44 – 104.3 = 0 dB   
2
cm
Table 2. Characteristic Magnetic Field Strength 
 
 
Input Power 
(incident) 
Cokeva 11940C  Operation Note
(
)
µA/m
 
µV
 = 60.3 dBµV.  Let K = 0 and AF = 44 dB
SA
mW
mW
 = 1 
2
2
cm
cm
 
Magnetic Field Strength 
0 dBm 
0 dBm 
0 dBm 
0 dBm 
0 dBm 
 
 
(
 
 
(H ),  dB  (µΑ/m) 
101 
98 
94 
93 
98 
 
)
µA/m
 
 
µV